Dissection of tissues in Endosurgery for obesity: EXPERIMENTAL ANALYSIS OF ligated

Cover Page

Full Text

Abstract

In the accomplished prospective research was fulfilled a comparison between ForceTriad LigaSure-V and Harmonic Scalpel in the circumstances of pretending intraabdominal environment according to lateral heat diffusion, ligation time, burst pressure, intensiveness forming high dispersive aerosol. On isolated vessels FT LS-V has shown the best burst pressure and the shortest ligation time. While using Harmonic Scalpel there was the least effect of lateral heat diffusion and the least smoke. In the state of obesity is held the reliability of the vessel ligation by increasing lateral heat diffusion and intensively of the smoke forming.

References

  1. Пучков К. В., Иванов В.В. Технология дозированного лигирующего электротермического воздействия на этапах лапароскопических операций, ИД Медпрактика-М. 2005, 176 с.
  2. Федоров И.В., Никитин А.Т. Клиническая электрохирургия. М., «Геотар Медицина», 1997
  3. Barrett WL, Garber SM. Surgical smoke - a review of the literature. Is this just a lot of hot air? Surg Endosc 2003, 17(6), 979-987
  4. Campbell PA, Cresswell AB, Frank TG, Cuschieri A Real-time thermography during energized vessel sealing and dissection/ Surg Endosc/ 2003,17(10) 1640-1645.
  5. Carbonell AM, Joels CS, Kercher KW, Matthews BD, Sing RF, Heniford ВТ: A comparison of laparoscopic bipolar vessel sealing devices in the hemostasis of small-, medium-, and large-sized arteries. J Laparoendosc Adv Surg Techniques 2003; 13(6):377-380.
  6. Clements RH, Palepu R: In vivo comparison of the coagulation capability of SonoSurg and Harmonic Ace on 4 mm and 5 mm arteries. Surg Endosc 2007;21(12):2203-2206.
  7. Hruby GW, Marruffo FC, Durak E, et al.: Evaluation of surgical energy devices for vessel sealing and peripheral energy spread in a porcine model. J Urol 2007;178(6): 2689-2693.
  8. Kim FJ, Chammas MF, Jr., Gewehr E, et al.: Temperature safety profile of laparoscopic devices: Harmonic ACE (ACE), Ligasure V (LV), and plasma trisector (PT). Surg Endosc 2008;22(6):1464-1469.
  9. Landman J, Kerbl K, Rehman J, et al.: Evaluation of a vessel sealing system, bipolar electrosurgery, harmonic scalpel, titanium clips, endoscopic gastrointestinal anastomosis vascular staples and sutures for arterial and venous ligation in a porcine model. J Urol 2003;169(2):697-700.
  10. Lee SJ, Park KH: Ultrasonic energy in endoscopic surgery. Yonsei Med J 1999;40(6):545-549.
  11. Richter S, Kollmar O, Neunhoeffer E, Schilling MK, Menger MD, Pistorius G: Differential response 'of arteries and veins to bipolar vessel sealing: evaluation of a novel reusable device. J Laparoendosc Adv Surg Techniques 2006;16(2):149-
  12. Weld KJ, Dryer S, Ames CD, et al.: Analysis of surgical smoke produced by various energy-based instruments and effect on laparoscopic visibility. J Endourol 2007;21(3): 347-351.

Supplementary files

Supplementary Files
Action
1. JATS XML

Copyright (c) 2010 Ivanov V.V., Puchkov K.V., Osipov A.V., Ivanov V.V., Puchkov K.V., Osipov A.V.

Creative Commons License
This work is licensed under a Creative Commons Attribution 4.0 International License.

Media Registry Entry of the Federal Service for Supervision of Communications, Information Technology and Mass Communications (Roskomnadzor) PI No. FS77-76803 dated September 24, 2019.



This website uses cookies

You consent to our cookies if you continue to use our website.

About Cookies