Implantation of a permanent left bundle branch pacing lead in a child after tetralogy of fallot repair. A case study

Cover Page


Cite item

Abstract

Conduction system pacing, including His bundle and left bundle branch pacing, is an increasingly utilized strategy in adult patients with bradyarrhythmias and atrioventricular conduction disorders. This technique helps prevent the development of heart failure caused by electrical and mechanical dyssynchrony associated with chronic ventricular pacing. These outcomes support the feasibiliy of this technique in younger pediatric patients under comparable clinical conditions. A distinct category includes children who have undergone congenital heart defect correction and require permanent cardiac pacing. These patients are at higher risk of developing pacing-induced сardiomyopathy due to non-physiologic ventricular contraction, while options for prevention and management remain limited. Biventricular pacing is technically challenging in pediatric patients, and algorithms for minimizing ventricular pacing cannot be applied in the presence of complete atrioventricular block. We report a case of left bundle branch pacing in a child with a history of surgical repair of tetralogy of Fallot and implantation of a permanent pacemaker for postoperative complete atrioventricular block.

Full Text

INTRODUCTION

Conduction system pacing has opened a new chapter in the use of implantable antiarrhythmic devices. In particular, implantation of a ventricular lead in the left bundle branch (LBB) area minimizes electrical dyssynchrony of the left ventricle, thereby preserving its physiological activation and contraction. This technique significantly reduces the risk of pacing-induced cardiomyopathy, demonstrating its advantages, particularly in patients with septal right ventricular pacing. In some cases, it also allows permanent cardiac pacing to be considered as an alternative to cardiac resynchronization therapy [1, 2]. The successful experience with left bundle branch pacing (LBBP) in patients older than 18 years suggests the feasibility of this technique even in younger children [3]. A specific patient category includes children who have undergone congenital heart defect (CHD) correction and require permanent cardiac pacing due to incisional atrioventricular block and other conduction disturbances in the postoperative period. The most extensive experience with ventricular lead implantation in patients with CHD has been accumulated with regard to correction of perimembranous ventricular septal defects. However, there are also sporadic reports on the use of conduction system pacing (CSP) in complex congenital defects, such as tetralogy of Fallot, double outlet right ventricle, atrioventricular canal, and others [4, 5].

The aim of this study is to describe a clinical case of ventricular lead implantation in the left bundle branch (LBB) in a child who previously underwent radical correction of tetralogy of Fallot.

CASE DESCRIPTION

At admission to Morozovskaya Children’s City Clinical Hospital in Moscow, the child was 9 years old, weighed 26 kg, and had a body surface area of 0.96 m².

At the age of 9 months, the child underwent radical correction of tetralogy of Fallot. The early postoperative period was complicated by complete atrioventricular block, requiring the implantation of a pacemaker system with an epicardial lead to the left ventricle. For 8 years, the patient was on continuous single-chamber ventricular pacing in VVIR mode, then a pacemaker replacement was carried out due to battery depletion. During routine testing, ventricular lead fracture and dysfunction were detected, prompting the recommendation to replace the epicardial system with an endocardial dual-chamber system, including ventricular lead implantation into the left bundle branch (LBB).

The preoperative electrocardiogram (ECG) showed epicardial left ventricular pacing in VVI mode with a QRS complex duration of 152 ms (Fig. 1a). According to the Echo data, left ventricular contractile function was normal prior to surgery.

 

Fig. 1. Twelve-lead electrocardiogram before and after pacemaker system replacement: (a) single-chamber ventricular pacing via an epicardial ventricular lead; (b) sequential dual-chamber pacing with left bundle branch pacing. ST–V6 interval, 52 ms; V6V1 interval, 52 ms; QRS duration, 136 ms.

 

Using a specialized delivery system designed for targeted lead implantation in the His bundle region (Select Site C315 His, Medtronic, USA; 2.4 mm [7Fr] diameter), a stylet-free active fixation lead (Select Secure 3830, Medtronic, USA; 59 cm long) was screwed into the interventricular septum under fluoroscopic guidance, with monitoring of electrical impedance and stimulated QRS morphology. The impedance at the final lead placement site was 923 ohm; the R-wave amplitude ranged 9.6–11.3 mV; the pacing threshold was 0.5 V with a pulse duration of 0.40 ms. The QRS complex duration in lead V6 was 134 ms; the peak-to-peak interval (RV1–RV6) was 52 ms; the St–RV6 interval was 52 ms. The atrial lead was implanted in the right atrial appendage. Upon completion of the procedure, a multi-projection assessment of lead positioning was performed (Fig. 2). The pacemaker was programmed to DDD mode with a basic pacing rate of 70 bpm. Left bundle branch pacing (LBBP) enabled a more physiological pacing mode, reducing the QRS complex duration from 152 ms to 136 ms (see Fig. 1b).

 

Fig. 2. Chest radiography after implantation of an endocardial dual-chamber pacemaker system with a ventricular lead in the left bundle branch: (a) right oblique view; (b) anteroposterior view; (c) left oblique view. © Dishekov et al., 2024.

 

CONCLUSION

In some cases, the correction of certain congenital defects, such as tetralogy of Fallot, may require the implantation of a permanent pacemaker, making patients lifelong-dependent on permanent cardiac pacing. The lead implantation site largely determines the prognosis and the risk of complications. For example, it is now well established that right ventricular apical lead placement is strongly associated with a high risk of intraventricular and interventricular electrical and mechanical dyssynchrony, which may ultimately contribute to the development of pacing-induced cardiomyopathy. Accumulating experience with His bundle and bundle branch pacing indicates a more favorable prognosis and the achievement of more physiological conduction, contributing to an increase in conduction system pacing procedures.

ADDITIONAL INFORMATION

Authors’ contribution. All the authors participated in the clinical case and the preparation of the article, read and approved the final version before publication. M.R. Dishekov, study design and concept, text writing, pacemaker implantation; M.V. Gorev, participation in the pacemaker implantation procedure, text edit, preparation for the publication; E.A. Talalaeva, participation in the pacemaker implantation procedure, text writing, scientific consulation; M.A. Abramyan, study design and concept, text edit.

Consent for publication. The authors have received written informed voluntary consent from the patient’s legal representatives to publish personal data.

Competing interests. The authors declare that they have no conflict of interest.

Funding source. This article was not supported by any external sources of funding.

×

About the authors

Murat R. Dishekov

Morozovskaya Children’s City Clinical Hospital

Email: mdishekov@gmail.com
ORCID iD: 0000-0002-1395-7827
SPIN-code: 3442-9013

cardiovascular surgeon, MD, Cand. Sci. (Medicine)

Russian Federation, Moscow

Maxim V. Gorev

City Clinical Hospital No 52, Moscow

Author for correspondence.
Email: drgorevmv@gmail.com
ORCID iD: 0000-0003-1300-4986
SPIN-code: 3572-2389

electrophysiologist

Russian Federation, Moscow

Elena A. Talalaeva

Morozovskaya Children’s City Clinical Hospital

Email: lealta27@gmail.com
ORCID iD: 0000-0001-5476-344X
SPIN-code: 6583-4970

cardiovascular surgeon

Russian Federation, Moscow

Mikhail A. Abramyan

Morozovskaya Children’s City Clinical Hospital

Email: mabramyan@morozdgkb.ru
ORCID iD: 0000-0003-4018-6287
SPIN-code: 4299-1032

MD, Dr. Sci. (Medicine)

Russian Federation, Moscow

References

  1. Li J, Jiang H, Cui J, et al. Comparison of ventricular synchrony in children with left bundle branch area pacing and right ventricular septal pacing. Cardiol Young. 2023;33(10):2078–2086. doi: 10.1017/S1047951122003675 EDN: WARJNT
  2. Moore JP, de Groot NMS, O'Connor M, et al. Conduction system pacing versus conventional cardiac resynchronization therapy in congenital heart disease. JACC Clin Electrophysiol. 2023;9(3):385–393. doi: 10.1016/j.jacep.2022.10.012 EDN: LJFSTJ
  3. Li J, Jiang H, Zhang Y, et al. A study to analyse the feasibility and effectiveness of left bundle branch area pacing used in young children. Pediatr Cardiol. 2024;45(3):681–689. doi: 10.1007/s00246-023-03119-8 EDN: UGHYEZ
  4. Cano Ó, Moore JP. Conduction system pacing in children and congenital heart disease. Arrhythm Electrophysiol Rev. 2024;13:e19. doi: 10.15420/aer.2024.09
  5. Chubb H, Mah D, Dubin AM, et al. Conduction system pacing in pediatric and congenital heart disease. Front Physiol. 2023;14:1154629. doi: 10.3389/fphys.2023.1154629 EDN: UZVQHV

Supplementary files

Supplementary Files
Action
1. JATS XML
2. Fig. 1. Twelve-lead electrocardiogram before and after pacemaker system replacement: (a) single-chamber ventricular pacing via an epicardial ventricular lead; (b) sequential dual-chamber pacing with left bundle branch pacing. ST–V6 interval, 52 ms; V6–V1 interval, 52 ms; QRS duration, 136 ms.

Download (420KB)
3. Fig. 2. Chest radiography after implantation of an endocardial dual-chamber pacemaker system with a ventricular lead in the left bundle branch: (a) right oblique view; (b) anteroposterior view; (c) left oblique view. © Dishekov et al., 2024.

Download (299KB)

Copyright (c) 2024 Eco-Vector

Creative Commons License
This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.

СМИ зарегистрировано Федеральной службой по надзору в сфере связи, информационных технологий и массовых коммуникаций (Роскомнадзор).
Регистрационный номер и дата принятия решения о регистрации СМИ: серия ПИ № ФС 77 - 79865 от 18.12.2020.